YY/T0573.2-2018,GB9706.27-2015 syringe flow tester
Instrument characteristics
This equipment is the testing equipment to test the product performance of the manufacturer.Composed of programmable controller, touch screen, force value sensor, transmission device, airborne printer, etc., it provides menu display in Chinese and English, with automatic operation test mode for various parameters of man-machine dialogue setting.In addition, the production unit, production batch machine input, so that the test results more standardized.Information security adopts three-level password to make information security and reliable.Test control range includes: nominal capacity, core rod displacement, printing setting, testing, uplink, downlink, time, calibration, verification and other function setting.The print test can display test load, sliding data (starting force, average force, zui high thrust) in real time.To automatically judge the test results, the automatic beep alarm is activated, and the printer will print out the test data and the results for analysis.
Test principle:
Measure the flow characteristics of the syringe (i. e. the maximum time taken to achieve the steady flow, the total deviation of the flow / setting output flow rate)
technical parameter:
Operation interface: Simplified Chinese / English
Test criteria: GB 15810-2019, YY / T 0573.2-2018, GB9706.27-2015, ISO7886-1:2017
Test item requirements: Set the drive to a flow rate equivalent to the I m L/h of the tested syringe.Set the computer sampling interval to O.5min.Open the drive to start the test.The test lasted 2H
The flow rate is calculated at 0.5 min intervals, and a solid flow rate line is drawn on the flow rate (mL / h) to time (min) curve chart
Test rate: 0-100mm / min Error range: ± 1mm / min; (the test speed can be set by software)
Nominal capacity: 0.1ml~200 ml, all specifications of arbitrary input.
Product net weight: 15kg
Overall Dimensions: 430290400 (mm)
Power supply: AC90V-240V / 50Hz (adaptive wide voltage)
Complete machine power: 70W
Programmable controller: PLC / ARM
Touch screen: 7-inch nylon pass color touch screen K600 + kernel / dot matrix 65 color 800 * 480 resolution
Sensor: High-precision force value sensor
Force measuring mechanism: built-in
Printer: an airborne needle printer
Equipment application:
The syringe flow characteristic meter is mainly used to determine the mobility energy physical characteristics of medical syringes, which is one of the important means to identify medical syringes.