Melt flow rate test

    Traffic : 0
    Edit time : 2019-08-12
    Relevant tags : 浦东新区Universal abrasion tester

    Application: melt flow rate test this instrument is not only suitable for testing polyethylene, polypropylene, polystyrene, ABS resin and other plastics with lower melting temperature, but also for testing polycarbonate, fluoroplastics, nylon and other engineering plastics with higher melting temperature

    Melt flow rate test standard:

    According to GB/T 3682-2000 thermoplastic melt mass flow rate and melt volume flow rate Determination, ISO 1133-2005 Plastics-Determination of the melt mass-flow rate(MFR) and the melt volume-flow rate(MVR) and other standards designed and manufactured for the Determination of thermoplastic melt flow rate instrument.

    Model csi-f686 is standard configuration, using quality test method, automatic scraping material (manual scraping material can also be used when the instrument is adjusted), the spline weight can be recorded into the instrument system to automatically calculate the results, and the test records can be saved for future reference, at the same time, it is equipped with a full set of weights and all relevant accessories (electronic scale is optional). Based on csi-f686, csi-f686 is equipped with a miniature printing device, which can print any existing test records in the form of supermarket receipts with very detailed printing contents. On the basis of csi-f686 type, melt density test and volumetric test mode are added, which makes the function more powerful and advanced.



      

    image.png 

    User guide:

    If you are using this kind of instrument for the first time, you can directly consult our technical staff, who will guide you through the whole process (from the name and function of accessories to all steps of the test, and then to the cleaning and maintenance of the instrument and matters needing attention). Out after opening the instrument, and its related accessories on the test bench, the anchor installation, by adjusting the anchor to instrument level state, the mouth will die into the material at the bottom of the cone zui, set temperature and test related parameters, after reaching the temperature of the set and keep constant temperature, the plastic particles under test through the funnel and feed screw feed tube, put the piston rod in again, will be stipulated by the load (weights) pressure on the piston rod, for piston rod drop to calibration, click on the "test" button start timing blanking, after the test, will be cut by the plastic spline weighing and saved to the instrument, the instrument will automatically calculation, and saved as a result, Convenient for future reference, at the same time, it can also print the recorded test records.

    Introduction to parameters:

    Adopt quality method to test; With automatic cutting device; With melt index calculation function, with FLASH memory, used to store test results. The instrument is simple in structure, convenient in use, accurate in measurement and reliable in performance. This instrument is not only suitable for polyethylene, polypropylene, polystyrene, ABS resin with lower melting temperature, such as plastic test, can also be applied to high temperature of polycarbonate, fluorine plastic, nylon engineering plastics such as testing, therefore is widely used in plastic production, plastic products, petrochemical and other industries as well as the relevant universities and colleges, scientific research units, the commodity inspection department.



      

    image.png 

    Instrument features:

    1. LCD Chinese display, SMD keyboard input, time, temperature and sampling times can be set intuitively and conveniently.

    2, the use of 32-bit high-end single-chip microcomputer as the control core, fast computing speed, high data accuracy.

    3. The 24-bit AD analog-to-digital converter is adopted in the United States, which provides a physical guarantee for the accuracy of temperature control.

    4. Embedded operating system is adopted to ensure reasonable internal task scheduling, accurate control of timing and delay, and accuracy of temperature control, sound prompt and other operations.

    5. Test results can be calculated, saved and printed.

    6. Adopt quality method (MFR) test.

    7. Cut the material automatically during the test and manually when the instrument is adjusted.

    8. Beautiful appearance, reasonable design, stable performance and convenient operation.

    Main functions:

    1. Melt flow rate test function of mass method.

    2. Automatic cutting function.

    3, mass melting index calculation function.

    4, melt index test results save function.

    5. Check, delete and print test records.

    Technical parameters:

    1. Extrude the discharging part

    Discharging mouth diameter: Φ 2.095 + / - 0.005 mm

    Discharge port length: 8.000±0.025 mm

    Charging tube inner diameter: Φ 9.550 + / - 0.025 mm

    Cartridge length: 152±0.1 mm

    The piston rod head diameter: Φ 9.475 + / - 0.015 mm

    Piston rod head length: 6.350±0.100 mm

    Piston rod effective stroke: 30mm

    2. Test load

    The test load is obtained by combining different amounts of weights, which can form the nominal load required by the test. There are 8 levels:

    Level 1:0.325kg=(piston rod + weight tray + heat shield +1# weight body)= 3.187n

    Level 2:1.200kg(0.325kg+2#0.875kg weight)=11.77N

    Grade 3:2.160kg=(0.325kg+3#1.835kg weight)=21.18N

    Grade 4:3.800kg=(0.325kg+4#3.475kg weight)= 37.26n

    Class 5:5.000kg=(0.325kg+5#4.675kg weight)= 49.03n

    Grade 6:10.000kg=(0.325kg+5#4.675k weight +6#5.000kg weight)=98.07N

    Grade 7:12.500kg=(0.325kg+5#4.675kg weight +6#5.000kg weight +7#2.500kg weight)= 122.58n

    Grade 8:21.600 kg=(0.325kg+2#0.875kg weight +3#1.835kg weight +4#3.475kg weight +5#4.675kg weight +6#5.000kg weight +7#2.500kg weight +8#2.915kg weight)= 211.82n

    3. Temperature range: room temperature ~ 400℃

    4. Temperature control accuracy: ±0.2℃

    5. Temperature fluctuation: ±0.2℃

    6. Temperature resolution: 0.1℃

    7. Temperature recovery time: ≤4min

    8. Test method: quality method (MFR)

    9. Measuring range: o. 1 ~ 300g/10min(MFR)

    10. Timing accuracy: 0.1s (MFR, MVR)

    11. Cutting method: automatic (during test), manual (during instrument adjustment)

    12. Zui large segment cutting times: 9 times

    13. Zui large cutting interval time: 9999.9 seconds

    14. Results: 20 copies

    15. Power supply: 220V, 50Hz

    16. Power: 490W

    17. Overall dimensions: long 290 x thick 410 x 630(mm)

    18. Printer configuration: panel thermal printer



    Shanghai chengsi intelligent technology co., LTD

    Contact: Mr. Gao

    Phone: 17701637746

    Phone: 19821311892

    Address: no. 265, east street, zhuangxing town, fengxian district, Shanghai