Shanghai chengsi intelligent technology co., LTD
Contact: Mr. Gao
Phone: 17701637746
Phone: 19821311892
Address: no. 265, east street, zhuangxing town, fengxian district, Shanghai
Pneumatic and sealing performance tester for abdominal puncture device
1 Description of the instrument application
YY / T1710-2020 is used to detect medical instruments such as disposable abdominal puncture air resistance and seal performance test.
2. System configuration requirements
2.1 One instrument main machine
2.2 One onboard printer
2.3 Supporting tooling
3 Technical parameters
3.1 Pressure range: higher than the local atmospheric pressure, 0~150 kPa, LED number display, the accuracy of ± 0.1%F.S or better; the pressure value can be set arbitrarily
3.2 There is inflation time, detection time and exhaust time, the range of 0 ~120 seconds or wider, adjustable, LED display, the error is not greater than ± 0.5S
3.3 PLC programmable controller, special HY LCD touch screen
3.4 Flow measurement range: 0 ~ 5 L/min or wider
3.5 Flow resolution: 1 mL/min
3.6 Complete test tooling 3 / 5 / 10 / 12 / 15 is required
specifications | size (mm) | specifications | size (mm) |
3 (Metal) | 3.5±0.3 | 3 (Plastic) | 4.2±0.3 |
5 (Metal) | 5.9±0.3 | 5 (Plastic) | 6.4±0.3 |
10 | 12.7±0.3 | 12 | 14.5±0.3 |
15 | 17.6±0.3 |
4 Acceptance indicators
4.1 Pressure range: higher than the local atmospheric pressure, 0~150 kPa, LED number display, the accuracy of ± 0.1%F.S or better; any pressure value can be set;
.24. Air-filling time, detection time and exhaust time, the range of 0 ~120 seconds or wider, adjustable, LED display, the error is not greater than ± 0.5S;
4.3 Flow measurement range: 0 ~ 5 L/min;
4.4 Flow resolution: 1 mL/min.
Shanghai chengsi intelligent technology co., LTD
Contact: Mr. Gao
Phone: 17701637746
Phone: 19821311892
Address: no. 265, east street, zhuangxing town, fengxian district, Shanghai