Curtain combustion tester

    Traffic : 0
    Edit time : 2019-08-12
    Relevant tags : 崇明Universal abrasion tester

    Curtain combustion tester

    Instrument is introduced

    The test principle of

    The fabric film flame propagation tester (drape-burning tester) is developed in accordance with NFPA 701-1 and is suitable for testing flame spread by burning flame from sample surface or bottom edge without external radiation.

    A dowel is used to hold the moistened sample vertically in the sample combustion chamber and a standard combustion lamp is used to light 45s. Observe the imagination of measuring combustion, measuring the time of combustion and weight changes before and after combustion.

    Test standard

    This fabric film flame propagation tester meets but is not limited to the following standards:

    NFPA 701-1

    Technical specifications


      

    image.png 

    Instrument size: 800 × 750 × 630 mm;

      

    image.png 

    Cavity size: 700 mm 750 mm 460 mm;

      

    image.png

    Air inlet: 7.0mm;

      

    image.png

    Inlet pressure: ≤1 MPa

      

    image.png

    Burner: meckel (fischer) blowtorch

      

    image.png

    Igniter: electric arc

      

    image.png 

    Weight: 40 kg;

      

    image.png 

      

    Power supply: 220 V

    The test object

    This instrument is developed for NFPA701 method 1, and the test objects include curtain, stage curtain, curtain, textile screen, tablecloth, etc. Thicker or coated materials are not suitable for this method.

    Composition and use of instrument

    listing

    This instrument should be carefully unpacked and inspected. For any damage or loss of the instrument, please contact the customer service department of xi company.

    When you open the package, please check the following materials (subject to the actual contract) :


      

    image.png

    1 main engine;

      

    image.png

    One shaft pin (including nails) for fixing the sample;

      

    image.png

    Shaft pin fasteners for fixing samples;

      

    image.png

    100mm(90mm~110mm) flame ruler

    Observation window (combustion chamber) : open the front door of the instrument during test, and close the front door when not in use to keep the instrument clean.

    Air vent: air guide fan is set;

    Timer 1, timing button 1: precision 0.1s, press the timing button 1, start the timer 1, press the timing stop again, press the yellow button on the timer to clear the timer;

    Timer 2, timing button 2: : precision 0.1s, press the timing button 1, start the timer 1, again press the timing stop, press the yellow button on the timer to clear the timer;

    Power switch: main power switch, indicator light means power on;

    Intake pressure gauge: shows the gas pressure entering the instrument;

    Pressure adjustment knob: the external gas will be reduced to the instrument can be used gas;

    Flow regulation: regulates the flow of gas into the burner.


      

    image.png 

    Lamps and lanterns to adjust

    This instrument USES imported merkel (fischer) blowtorch, suitable for all types of gas, equipped with damper and flame adjusting knob.

    The tip of the ignition needle shall be installed on the surface of the blowtorch, about 5mm from the edge and above the vertical distance of about 5mm.

    After the flame is lit, adjust the flow rate on the adjustment panel to make the flame size moderate, and adjust the damper and the flame fine tuning knob to adjust the color of the flame. When methane or other gases are used as gas, this test requires a flame height of 100mm and a blue flame.

    Sampling and preparation

    Sampling shall be performed in accordance with method 1 in standard NFPA 701, with appropriate fine-tuning if non-standard tests are required.

    The sample size is 150mm×400mm, and the sample is cut along the longitude and latitude direction. The length direction is the vertical direction, and the vertical direction of the actual use position of the material is the length direction. For multilayer sample measurements, please refer to NFPA 701 for suture treatment.

    The sample should be dried at 105℃ for at least 30min. If the sample will undergo deformation or other reactions at 105℃, the sample should be dried at 20℃ for at least 24h.

    A minimum of 10 samples are prepared from the standard test, each of which is numbered and weighed before combustion to 0.001g(or as otherwise required).

    The sample is then placed on the internal sample rack of the instrument. The sample bracket can be adjusted by means of screws on both sides to make the lower end of the sample flush with the flat burner center. The position of the sample rack is shown in the figure. The height and height screws are adjusted according to the figure:

    Flame adjustment

    Confirm that after the blowtorch is laid flat, the center of the blowtorch head is 25mm away from the sample (the instrument has been adjusted when leaving the factory).

    Keep gas lamp upright and connect gas. Make sure that the air pressure adjusting knob, flow adjusting knob and blowtorch adjusting knob are properly opened. Press start button to ignite the blowtorch. If not, please check whether the air path is smooth. Adjust the flame until the yellow flame disappears, showing orange or blue flame, and ensure the flame height is 100mm.

    Keep the flame steady for at least 5 minutes. A flame suitable for testing is shown below:


      

    image.png 

    Fire test

    Place the burner upside down and push the burner into the sample position. The flame starts to ignite the sample. After 45s, the system cuts off the gas and the torch stops burning automatically.

    Observe whether the sample starts to burn. If the sample burns, press timer 1 to start the timer, and stop the timing after the sample burns.

    At the same time, observe whether there is any droplet. If there is any droplet combustion, press timer 1 for timing, and stop timing after the sample combustion. Afterburning time was revised to the maximum duration of sample and low content combustion.



    Shanghai chengsi intelligent technology co., LTD

    Contact: Mr. Gao

    Phone: 17701637746

    Phone: 19821311892

    Address: no. 265, east street, zhuangxing town, fengxian district, Shanghai

    备案号:沪ICP备19009154号-12